Chenfan Electric
EN
Send drawing

ARTICLETechnical article

Helium Leak Testing of Transformer Tanks: Background and Recovery

  • Chenfan Power

A sensitive helium detector can respond to tracer left by an earlier test or present around its inlet. A small reading is not automatically a tank leak, while subtracting a drifting background can also conceal a real one. Preserve the signal history and the test boundary. [1]

Helium Leak Testing of Transformer Tanks: Background and Recovery: technical review table
First article table, reproduced without adding or changing values. See the article for assumptions and application boundaries.

Read the trace in four parts

  1. Background: record the stable or changing baseline before the defined tank test.

  2. Reference response: verify the instrument through the approved calibration/check arrangement.

  3. Test response: associate each reading with location, time and the applied tracer boundary.

  4. Recovery: record return toward baseline and any carry-over before another location or tank is tested.

NASA’s tracer-gas guidance couples measured background with setup calibration and calls for further quantitative investigation of suspect probe indications. [1] For the tank, the approved procedure defines pressure, tracer application, detector operation and recovery checks.

Track the source of background

Possible influenceEvidenceDiagnostic question
Previous test tracerTime and location historyDoes signal persist away from the object?
Ambient air movementVentilation and inlet positionDoes transport change response timing?
Instrument recoveryBaseline and reference behaviorIs carry-over affecting the next measurement?
True test pathRepeatable localized response under defined conditionsDoes independent evidence support the route?

Combine standard uncertainties with covariance where effects are correlated; repeatability alone does not include all systematic contributions. [2]

Bound a background-corrected leak indication

Assumptions at 20 °C under the same approved test configuration: test and representative background readings are 8×10⁻⁸ and 3×10⁻⁸ Pa·m³/s, with independent standard uncertainties 1×10⁻⁸ and 2×10⁻⁸ Pa·m³/s. Corrected indication is 5×10⁻⁸ Pa·m³/s and uc = 2.236×10⁻⁸ Pa·m³/s. With assumed coverage factor 2, its interval is 0.528–9.472×10⁻⁸ Pa·m³/s. An assumed tank criterion requires the upper decision value to be at most 8×10⁻⁸ Pa·m³/s; the present result cannot demonstrate compliance, despite its lower nominal value. Hold release and resolve background stability/recovery and the localized path before retesting. This interval by itself does not prove an excessive tank leak. [2]

References

  1. NASA, NASA-STD-7012A, Leak Test Requirements, 2023, sections 4.2 and 5.11.
  2. B. N. Taylor and C. E. Kuyatt, NIST Technical Note 1297, Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results, 1994, sections 5 and 6.
Chenfan Electric

Direct coordination

Start a conversation.

A drawing, a material requirement, or a project to discuss.

Send drawing