Chenfan Electric
EN
Send drawing

ARTICLETechnical article

Quality Factor and Internal Voltage in Resonant HV Tests

  • Chenfan Power

Quality factor sets the drive required to produce object voltage and the load sensitivity of a resonant HV test circuit. Calculate internal component voltages and currents at that same operating point to establish the system duty.

HIGHVOLT describes low feeding power and a high-Q operating principle for its resonant system. The relevant loss boundary and component stresses still need to be stated before that description becomes an engineering calculation. [1]

Quality Factor and Internal Voltage in Resonant HV Tests: technical review table
Define what is included in Q

Define what is included in Q

For an ideal series RLC circuit at resonance, Q = ωL/R = 1/(ωCR), where R represents the equivalent series losses within the chosen boundary. With driving voltage Vdrive and current I, Vdrive = IR, while the magnitude of the capacitor voltage is VC = I/(ωC) = QVdrive.

The reactor’s inductive voltage magnitude is also approximately QVdrive in this ideal model. Its phase opposes the capacitor voltage, so the large component voltages cancel in the series sum; they do not vanish as insulation duties.

QuantityBoundary to identify
Driving voltageInput to the defined series test circuit
Object voltageVoltage across the actual capacitive test object
Reactor voltageVoltage across the reactor and its insulation system
Equivalent lossReactor, object and other included losses
Quality factorThe same stated frequency, load and loss boundary

Compare loaded and unloaded conditions carefully

An empty system does not necessarily have the same resonance or Q as the loaded circuit. Added object capacitance changes frequency in a fixed-inductance system, while dielectric losses and accessories change the effective loss.

If the object becomes more lossy, the driving voltage and real feeding power required to maintain its voltage can rise. A Q measured under a low-loss reference load therefore cannot establish the source burden for every object.

At resonance, real input power in the simple model is I²R. Low real feeding power can coexist with significant electric and magnetic stored energy. It should not be used to infer negligible fault exposure.

Keep amplification and sensitivity together

Higher Q narrows the resonance response in the ideal circuit. Frequency or capacitance changes can then materially change output for a fixed drive. The control system’s tuning and regulation behavior becomes part of the operating assessment.

Compare measured object voltage with the calculated component stresses and the supported control region. Near discharge onset or a fault, the linear loss model may cease to represent the circuit. Do not extend the steady-state Q relationship through that transient without further analysis.

Use Q to explain drive requirements and load sensitivity at the defined point. Verify component voltage, current, duration and protection duties before accepting that operating combination.

Calculate drive and internal stress at the same resonant point

Calculate capacitive current and resonance from the same circuit basis [2].

Assume a series-resonant 50 Hz circuit with C = 10 nF, object voltage 100 kV RMS and loaded Q = 50. Resonant inductance is 1,013.21 H, equivalent series resistance is 6.366 kΩ and current is 0.3142 A. Drive is 2.00 kV while real input loss is 628.3 W. The ideal reactor still carries 100 kV RMS. Rate each component at its own port and retain the stored-energy/fault assessment.

References

  1. HIGHVOLT Prüftechnik Dresden GmbH, Applied Voltage Test System for Transformers, 8.72/4 Type WRV M, November 2020.
  2. U.S. Department of Energy, DOE-HDBK-1011/3-92, Electrical Science, Volume 3, June 1992.
Chenfan Electric

Direct coordination

Start a conversation.

A drawing, a material requirement, or a project to discuss.

Send drawing